Related Experiment Video
Updated: Apr 19, 2026

Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Stent distortion after sutureless aortic valve implantation: a new complication seen with a novel surgical technique
Felix Fleissner1, Ulrich Molitoris2, Malakh Shrestha3
1Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany fleissner.felix@mh-hannover.de.
Abstract:
In recent years, sutureless aortic valves have grown in popularity. As these valves do not need to be anchored with sutures, shorter cardiac ischaemia and extracorporeal circulation times can potentially be achieved. In addition, the absence of a sewing ring in these valves results in a greater effective orifice area for any given size. Postoperative outcome may be improved using sutureless valves especially when combined with minimally invasive surgical techniques. However, sutureless aortic valves present unique surgical risks. We report 2 cases of delayed distortion of a sutureless aortic valve stent resulting in paravalvular leakage and increased transvalvular gradients. One patient underwent a reoperation with an aortic valve replacement using a standard biological aortic valve, the other patient was treated with balloon dilatation of the aortic valve stent.