Peri-operative pulse oximetry in low-income countries: a cost-effectiveness analysis
Samantha L Burn1, Peter J Chilton1, Atul A Gawande2
1School of Health and Population Sciences, University of Birmingham, Edgbaston, Birmingham B15 2TT, England .
Objective:
To evaluate the cost-effectiveness of pulse oximetry--compared with no peri-operative monitoring--during surgery in low-income countries.
Methods:
We considered the use of tabletop and portable, hand-held pulse oximeters among patients of any age undergoing major surgery in low-income countries. From earlier studies we obtained baseline mortality and the effectiveness of pulse oximeters to reduce mortality. We considered the direct costs of purchasing and maintaining pulse oximeters as well as the cost of supplementary oxygen used to treat hypoxic episodes identified by oximetry. Health benefits were measured in disability-adjusted life-years (DALYs) averted and benefits and costs were both discounted at 3% per year. We used recommended cost-effectiveness thresholds--both absolute and relative to gross domestic product (GDP) per capita--to assess if pulse oximetry is a cost-effective health intervention. To test the robustness of our results we performed sensitivity analyses.
Findings:
In 2013 prices, tabletop and hand-held oximeters were found to have annual costs of 310 and 95 United States dollars (US$), respectively. Assuming the two types of oximeter have identical effectiveness, a single oximeter used for 22 procedures per week averted 0.83 DALYs per annum. The tabletop and hand-held oximeters cost US$ 374 and US$ 115 per DALY averted, respectively. For any country with a GDP per capita above US$ 677 the hand-held oximeter was found to be cost-effective if it prevented just 1.7% of anaesthetic-related deaths or 0.3% of peri-operative mortality.
Conclusion:
Pulse oximetry is a cost-effective intervention for low-income settings.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...