Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
Published on: April 27, 2016
The use of shear rate-diameter dose-response curves as an alternative to the flow-mediated dilation test
Brian Wu1, Daniel Credeur2, Simon Fryer3
1Keck School of Medicine, Los Angeles, CA, United States.
Abstract:
The brachial artery flow-mediated dilation test (FMD) is the non-invasive gold-standard used to test endothelial function. Reduced FMD precedes the development of atherosclerosis and provides an early marker for predicting future cardiovascular disease events. Although, this test is of high potential, it is somewhat limited by poor reproducibility. By utilizing hand warming and grip exercise combined with hierarchical linear modeling, shear rate-diameter dose-response curves may provide a novel and more accurate way to assess endothelial function in humans. Shear rate-diameter dose-response curves could potentially improve upon the traditional FMD measurement and serve as a superior clinical and research tool for assessing cardiovascular disease risk in a variety of populations. The current paper presents testable hypotheses and methodology for assessing the validity and reliability of an alternative to the current FMD test.