Pain-only complaint about cochlear implant device: A five-patient pediatric experience

Insights

Pain at the cochlear implant (CI) receiver-stimulator site can be the only sign of device failure. Revision surgery resolved pain in most pediatric cases, highlighting the need for a CI pain registry.

Area of Science:

  • Otolaryngology
  • Biomedical Engineering
  • Pediatric Audiology

Background:

  • Pediatric cochlear implant (CI) surgeries are common, yet complications like receiver-stimulator site pain are infrequently reported.
  • This case series examines five pediatric patients experiencing isolated pain at the CI receiver-stimulator site.

Observation:

  • Pain onset varied from 2 to 16 years post-implantation and was unresponsive to medical therapy.
  • Pain persisted even when the external device was off or not worn.
  • Four patients had bilateral implants, but pain localized to only one receiver-stimulator.

Findings:

  • Revision surgery successfully resolved pain in four out of five pediatric CI patients.
  • Two explanted devices showed hermeticity loss and insulation failure.
  • Post-revision, three of four patients experienced good outcomes with their replacement CI.

Implications:

  • Isolated pain can be the sole indicator of cochlear implant device failure in pediatric patients.
  • A standardized flowchart for managing CI-related pain and a global registry are recommended.
  • Further research is needed to understand the underlying mechanisms of CI-induced pain.