Updated: Apr 18, 2026

A Porcine Model of Acute Autologous Pulmonary Embolism
07:44

A Porcine Model of Acute Autologous Pulmonary Embolism

Published on: September 6, 2024

992

Retrieval of distorted pipeline embolic device using snare-loop.

Bartley Mitchell1, Liang-Der Jou2, Michel Mawad3

  • 1Department of Neurosurgery, Baylor College of Medicine, 1709 Dryden, Suite 750, Houston, TX 77030, USA.

Journal of Vascular and Interventional Neurology
|January 8, 2015
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Development of multidisciplinary, evidenced-based protocol recommendations and implementation strategies for anterior lumbar interbody fusion surgery following a literature review.

Medicine·2023
Same author

Investigating the Relationship between Ambulatory and Hospital Patient Experience Scores in a Neurosurgery Practice.

Healthcare (Basel, Switzerland)·2021
Same author

In Reply to the Letter to the Editor Regarding "Pediatric Moyamoya Presenting as a Subarachnoid Hemorrhage from a Ruptured Anterior Cerebral Artery Aneurysm".

World neurosurgery·2020
Same author

Pediatric Moyamoya Presenting as a Subarachnoid Hemorrhage from a Ruptured Anterior Cerebral Artery Aneurysm.

World neurosurgery·2019
Same author

In vitro digital subtraction angiographic evaluation of flow diverters in a patient-specific aneurysm.

Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences·2017
Same author

Evaluation of Pipeline Flex delivery system for the treatment of unruptured aneurysms.

Expert review of medical devices·2016

Malpositioned flow-diverter stents, like the Pipeline Embolic Device (PED), can be safely removed using endovascular snare-loops. This allows for successful retreatment of intracranial aneurysms, preventing potential complications.

Area of Science:

  • Endovascular neurosurgery
  • Interventional neuroradiology
  • Medical device technology

Background:

  • Flow-diverter stents are increasingly utilized for intracranial aneurysm treatment.
  • Increased use may lead to higher rates of stent malpositioning, distortion, or migration.
  • Complications from malpositioned stents pose risks for iatrogenic stroke.

Purpose of the Study:

  • To describe the successful endovascular removal of a deformed Pipeline Embolic Device (PED) stent.
  • To highlight the utility of snare-loop devices for retrieving malpositioned endovascular devices.
  • To emphasize the importance of contingency planning in stent-based aneurysm treatment.

Main Methods:

  • A case report detailing the conformational twisting and deformity of a PED stent during carotid artery placement.
Keywords:
IntracranialPipeline embolic deviceaneurysmretrievalsnare loop

More Related Videos

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
05:52

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model

Published on: November 4, 2021

3.5K
Simulator Training for Endovascular Neurosurgery
08:08

Simulator Training for Endovascular Neurosurgery

Published on: May 6, 2020

4.3K

Related Experiment Videos

Last Updated: Apr 18, 2026

A Porcine Model of Acute Autologous Pulmonary Embolism
07:44

A Porcine Model of Acute Autologous Pulmonary Embolism

Published on: September 6, 2024

992
Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
05:52

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model

Published on: November 4, 2021

3.5K
Simulator Training for Endovascular Neurosurgery
08:08

Simulator Training for Endovascular Neurosurgery

Published on: May 6, 2020

4.3K
  • Description of immediate stent retrieval using an endovascular snare-loop.
  • Subsequent successful treatment of the aneurysm with a second PED stent after removal of the deformed device.
  • Main Results:

    • The malformed PED stent was successfully removed using a snare-loop device.
    • The intracranial aneurysm was subsequently treated effectively with a second PED stent.
    • Demonstrates the feasibility and safety of endovascular retrieval for deformed flow-diverter stents.

    Conclusions:

    • Endovascular snare-loop devices are effective for removing malpositioned or deformed flow-diverter stents.
    • Having a retrieval strategy is critical for managing complications associated with stent placement.