Safety considerations for wireless delivery of continuous power to implanted medical devices
Abstract:
Wireless power systems for use with implants are referred to as transcutaneous energy transmission systems (TETS) and consist of an implanted secondary coil and an external primary coil along with supporting electronics. A TETS system could be used to power ventricular assist systems and eliminate driveline infections. There are both direct and indirect safety concerns that must be addressed when continuously transferring power through the skin. Direct safety concerns include thermal tissue damage caused by exposure to the electromagnetic fields, coil heating effects, and potential unwanted nerve stimulation. Indirect concerns are those caused by potential interference of the TETS system with other implanted devices. Wireless power systems are trending towards higher frequency operation. Understanding the limits for safe operation of a TETS system across a range of frequencies is important. A low frequency and a high frequency implementation are simulated to demonstrate the impact of this trend for a VAD application.