Related Experiment Video
Updated: Apr 18, 2026

Hemodynamic Precision in the Neonatal Intensive Care Unit using Targeted Neonatal Echocardiography
Published on: January 27, 2023
Initial clinical experience with a miniaturized transesophageal echocardiography probe in a cardiac intensive care
Nick Fletcher1, Martin Geisen2, Hanif Meeran1
1Department of Intensive Care Medicine, St. Georges Healthcare NHS Trust, London, United Kingdom; Department of Anaesthesia, St. Georges Healthcare NHS Trust, London, United Kingdom.
Objective:
To investigate the safety of a novel, miniaturized, monoplane transesophageal echocardiography probe (mTEE) and its potential as a hemodynamic monitoring tool.
Design:
This was a retrospective analysis of the clinical evaluation of a disposable mTEE in ventilated patients with severe cardiogenic shock requiring hemodynamic support. mTEE assessment was performed by operators with mixed levels of TEE training. Information on hemodynamic interventions based on mTEE findings was recorded.
Setting:
A tertiary university cardiac critical care unit.
Participants:
Male and female critical care patients admitted to the unit with severe hemodynamic instability.
Interventions:
Insertion of miniaturized disposable TEE probe and hemodynamic and other critical care interventions based on this and conventional monitoring.
Measurements And Main Results:
In 41 patients (51.2% female, 73.2% after cardiac surgery), hemodynamic support probe insertion was accomplished without major complications. A total of 195 mTEE studies were performed, resulting in changes in therapy in 37 (90.2%) patients based on mTEE findings, leading to an improvement in hemodynamic parameters in 33 (80.5%) patients. Right ventricular (RV) failure was diagnosed in 25 patients (67.6%) and mTEE had a direct therapeutic impact on management of RV failure in 17 patients (68 %).
Conclusions:
Insertion and operation of a novel, miniaturized transoesophageal echocardiography probe can be performed for up to 72 hours without major complications. Repeated assessment using this device provides complementary information to invasive monitoring in the majority of patients and has an impact on hemodynamic management.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Cardiac Catheterization II: Right Heart Catheterization
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization