Computed tomography scanogram compared to long leg radiograph for determining axial knee alignment
Related Experiment Video
Updated: Apr 18, 2026

Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
Thomas J Holme1, Johann Henckel, Kai Hartshorn
1Department of Trauma & Orthopaedics, University College London.
Background And Purpose:
Supine computed tomography scanogram (CTS) is a commonly used alternative to weight bearing long leg plain radiograph (LLR) in measuring knee alignment. No published studies have validated its use in the native knee and the post-unicompartmental replacement knee (UKR). We quantified the difference in measurements obtained from CTS and LLR for knee alignment.
Patients And Methods:
Supine CT scanograms and weight bearing long leg plain anteroposterior radiographs were obtained for 40 knees (in 25 patients), 17 of which were native, and 23 of which were post-UKR. The mechanical and anatomical axes of the tibio-femoral joint were measured. Bland-Altman plots were used to calculate the 1.96 standard deviation limits of agreement between CTS and LLR. Intraclass correlation was used to assess intra-rater and inter-rater reliability (where values > 0.81 indicate very good reliability).
Results:
CTS and LLR were equally reliable in measurement of the mechanical and anatomical axes of the tibio-femoral joint (intraclass correlation coefficient (ICC) > 0.9 for all parameters). Statistically significant and clinically relevant differences were found between CTS and LLR in measurement of the mechanical axis (limits of agreement: UKR -3.2° to 6.3°; native -3.2° to 5.6°) and the anatomical axis (limits of agreement: UKR -3.7° to 8.7°; native -2.0° to 8.8°).
Interpretation:
Although it is a reliable tool, CTS is not necessarily an accurate one for measurement of knee alignment when compared to LLR. We recommend that CTS should not be used as a substitute for LLR in measurement of the mechanical or anatomical axes of the knee.
More Related Videos
07:43In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT