Related Experiment Video
Updated: Apr 18, 2026

Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
Published on: January 30, 2017
Evaluation of novel second-generation RSV and influenza rapid tests at the point of care
Ryan Tuttle1, Anja Weick1, Wiebke Sabrina Schwarz1
1Division of Pneumonology-Immunology, Department of Pediatrics, Charité University Medical Center, Augustenburger Platz 1, 13353 Berlin, Germany.
Abstract:
Acute respiratory infections represent common pediatric emergencies. Infection control warrants immediate and accurate diagnoses. In the past, first-generation respiratory syncytial virus (RSV) rapid tests (strip tests) have shown suboptimal sensitivities. In 2013, the Food and Drug Administration licensed a second-generation RSV rapid test providing user-independent readouts (SOFIA™-RSV) using automated fluorescence assay technology known to yield superior results with influenza rapid testing. We are reporting the first point-of-care evaluation of the SOFIA™-RSV rapid test. In the Charité Influenza-Like Disease Cohort, 686 nasopharyngeal samples were tested in parallel with SOFIA™-RSV and SOFIA™-Influenza A+B. Compared to real-time PCR, SOFIA™-RSV sensitivities/specificities were 78.6%/93.9%, respectively (SOFIA™-Influenza A: 80.6%/99.3%). Performance was greatest in patients below 2 years of age with a test sensitivity of 81.8%. RSV sensitivities were highest (85%) in the first 2 days of illness and with nasopharyngeal compared to nasal swabs (P=0.055, McNemar's test). Second-generation RSV and influenza rapid testing provides highly accurate results facilitating timely patient cohortation and management.