Related Experiment Video
Updated: Apr 18, 2026

Fabricating Superhydrophobic Polymeric Materials for Biomedical Applications
Published on: August 28, 2015
Application of hydrophobic coatings in biodegradable devices
Juan Meng1, Haiyan Li2, Yaqin Gao3
1Shanghai MicroPort Medical (Group) Co., Ltd, Shanghai, China Med-X Research Institute, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
Background:
Degradation periods of biodegradable medical devices strongly affect their clinical performance and therefore special attention has been drawn to modulate their degradation rate.
Objective:
This paper presents an experimental study on the effect of hydrophobic coating on the degradation behavior of PLLA samples.
Methods:
PLLA films were coated with a thin layer of PCL, and a combination of scanning electron microscopy (SEM), Ubbelohde Viscometer Capillary, and chromatograph of gel permeation (GPC) was used to evaluate the morphology and molecular weight changes of samples during degradation. In addition, the mass loss of samples was also measured during the experiment.
Results:
PLLA samples with PCL coatings showed a slower degradation rate than those without PCL coatings, which indicated that PCL coatings could protect inside PLLA samples and slow down the degradation rate of PLLA samples.
Conclusions:
The results of the study suggest that hydrophobic coating on polymer materials is a useful approach to control the degradation of polymer medical device.