Evaluation of an oxygen mask-based capnometry device in subjects extubated after abdominal surgery
Related Experiment Video
Updated: Apr 18, 2026

Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Shunsuke Takaki1, Takahiro Mihara2, Kenji Mizutani3
1Department of Anesthesiology and Intensive Care Unit, Yokohama City University Hospital, Kanagawa, Japan. shunty5323@gmail.com.
Background:
For early detection of respiratory and hemodynamic changes during anesthesia, continuous end-tidal carbon dioxide concentration (PETCO2) is monitored by capnometry. However, the accuracy of CO2 monitoring during spontaneous breathing in extubated patients remains undetermined. Therefore, we aimed to compare P(ETCO2) measured by capnometry using an oxygen mask with a carbon dioxide sampling port (capnometry-type oxygen mask) and P(CO2) in extubated subjects who had undergone abdominal surgery. Furthermore, we investigated whether spontaneous deep breathing affected dissociation between P(aCO2) and P(ETCO2).
Methods:
Adult post-abdominal surgery subjects admitted to the ICU were enrolled in this study. After extubation, oxygen was supplied at 6 L/min using the capnometry-type oxygen mask. After 30 min of oxygen supply, P(aCO2) blood gas analysis was performed, and P(ETCO2) was measured under resting and deep-breathing conditions. For both resting and deep-breathing conditions, the correlation between P(aCO2) and P(ETCO2) was analyzed. Furthermore, bias, precision, and limits of agreement were calculated using the Bland-Altman method.
Results:
Twenty-five subjects (15 men, 10 women) with a mean age of 62 y (interquartile range of 57-76 y) and body mass index of 20-24 kg/m(2) were studied. The correlation (r) between P(aCO2) and P(ETCO2) under resting and deep-breathing conditions was 0.50 and 0.56, respectively. Compared with P(aCO2), the bias and limits of agreement were -12.6 (-20.6 to -4.6) for resting P(ETCO2) and -9.1 (-16.0 to -2.1) for deep-breathing P(ETCO2). The association between P(aCO2) and deep-breathing P(ETCO2) was significantly smaller compared with resting P(ETCO2) (P = .002).
Conclusions:
It is possible to measure the P(ETCO2) under varying breathing conditions with the capnometry-type oxygen mask in subjects receiving oxygen supplementation after extubation following upper abdominal surgery to determine whether they are properly ventilating. (ClinicalTrials.gov registration UMIN000011925.).
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Cardiopulmonary Resuscitation II: ACLS Airway Management