Related Experiment Video
Updated: Aug 9, 2026

Performing Behavioral Tasks in Subjects with Intracranial Electrodes
Published on: October 2, 2014
Invasive EEG in refractory epilepsy: insertion of subdural grids through linear craniectomy reduces complications and
C Raftopoulos1, G Vaz1, D Tassigny1
1Department of Neurosurgery, University Hospital St-Luc, Université Catholique de Louvain, 10, Avenue Hippocrate, 1200 Brussels, Belgium.
Objective:
To evaluate our technique of implanting subdural grids by linear craniectomy under computer-assisted navigation for invasive electroencephalography in medically refractory epilepsy.
Material And Method:
We report results from our first 38 consecutive patients with medically refractory epilepsy who underwent subdural grids implantation by linear craniectomy. For each case, a preoperative MRI was performed for navigation followed by a postoperative MRI for localization control of the intracranial electrode contacts. A linear skin incision, adapted to the depth and type of subdural electrode (strip or grid) and compatible with possible subsequent therapeutic surgery, was carried out. One or two linear craniectomies (maximal length 6cm, width 1cm) were then drilled with a bevel. The dura mater was incised under microscopic guidance to avoid opening the arachnoid. The required subdural electrodes were then slipped subdurally through each linear craniectomy (letter-box technique).
Results:
Forty-one invasive electroencephalographies were performed with 28 (68%) bilateral. For all invasive electroencephalographies, at least one subdural grid was implanted. Sixty-one subdural grids were implanted in total, 52 with 20 contacts and nine with 32 contacts. No cerebrospinal fluid leakage, no infection, no neurological deficit and no permanent complications were observed. Three subdural grids (5%) were not positioned exactly as planned but this had no consequence for the invasive electroencephalography analysis.
Conclusion:
The implantation of 61 consecutive subdural grids for invasive electroencephalography through linear craniectomies was associated with no transient or permanent complications in this population. This letter-box technique appears to be practical and safe without limiting explorative efficacy.
More Related Videos
09:00Investigating the Function of Deep Cortical and Subcortical Structures Using Stereotactic Electroencephalography: Lessons from the Anterior Cingulate Cortex
Published on: April 15, 2015
08:30Subdural Soft Electrocorticography ECoG Array Implantation and Long-Term Cortical Recording in Minipigs
Published on: March 31, 2023