Related Experiment Video
Updated: Apr 18, 2026

Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
Published on: May 2, 2025
Timer switch to convert suction apparatus for negative pressure wound therapy application
Surath Amarnath1, Mettu Rami Reddy1, Chayam Hanumantha Rao1
1Department of Orthopaedics, NRI Medical College and General Hospital, Chinnakakani, Guntur, Andhra Pradesh, India.
Background:
Negative pressure wound therapy (NPWT) is an established modality in the treatment of chronic wounds, open fractures, and post-operative wound problems. This method has not been widely used due to the high cost of equipment and consumables. This study demonstrates an indigenously developed apparatus which gives comparable results at a fraction of the cost. Readily available materials are used for the air-tight dressing.
Materials And Methods:
Equipment consists of suction apparatus with adjustable pressure valve set to a pressure 125-150 mmHg. An electronic timer switch with a sequential working time of 5 min and a standby time of 3 min provides the required intermittent negative pressure. Readily available materials such as polyvinyl alcohol sponge, suction drains and steridrapes were used to provide an air tight wound cover.
Results:
A total of 90 cases underwent 262 NPWT applications from 2009 to 2014. This series, comprised of 30 open fractures, 21 post-operative and 39 chronic wounds. The wound healing rate in our study was comparable to other published studies using NPWT.
Conclusion:
The addition of electronic timer switch will convert a suction apparatus into NPWT machine, and the results are equally effective compared to more expensive counter parts. The use of indigenous dressing materials reduces the cost significantly.
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Tracheostomy Suctioning II: Procedure
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...