Relative flow reserve derived from quantitative perfusion imaging may not outperform stress myocardial blood flow for
Wijnand J Stuijfzand1, Valtteri Uusitalo1, Tanja Kero1
1From the Departments of Cardiology (W.J.S., I.D., M.T.R., K.M.M., N.v.R., P.K.), Radiology and Nuclear Medicine (P.G.R., A.A.L, H.J.H., M.C.H.), Department of Epidemiology and Biostatistics (M.W.H.), VU University Medical Center, Amsterdam, The Netherlands; Turku PET Centre, Turku University Hospital and University of Turku, Turku, Finland (V.U., A.S., S.A.K, M.P., J.K.); and Department of Nuclear Medicine and PET, Institution of Radiology, Oncology and Radiation Science, Uppsala University, Uppsala, Sweden (T.K., J.S.).
Insights
Quantitative myocardial perfusion imaging can noninvasively estimate fractional flow reserve (FFR) using relative flow reserve (RFR). While feasible, RFR showed only a trend toward improved diagnostic accuracy for coronary artery disease detection.
Area of Science:
- Cardiovascular Imaging
- Noninvasive Diagnostics
- Coronary Artery Disease
Background:
- Quantitative myocardial perfusion imaging is crucial for diagnosing coronary artery disease.
- Noninvasive calculation of fractional flow reserve (FFR) is possible using this imaging.
- Relative flow reserve (RFR) quantifies blood flow in stenotic versus normal areas.
Purpose of the Study:
- To evaluate the diagnostic value of RFR in detecting significant coronary artery disease.
- To compare RFR with invasive FFR and other quantitative perfusion parameters.
Main Methods:
- Oxygen-15-labeled water cardiac positron emission tomography was used in 92 patients.
- Invasive coronary angiography and FFR were performed for comparison.
- Hyperemic myocardial blood flow (MBF), coronary flow reserve, and RFR were calculated.
Main Results:
- 41% of intermediate lesions were hemodynamically significant (FFR ≤0.80).
- RFR and other perfusion parameters were significantly lower in significant lesions (P<0.001).
- RFR showed a moderate correlation with FFR (r=0.54) and an AUC of 0.82.
Conclusions:
- Noninvasive RFR estimation via quantitative perfusion PET is feasible.
- RFR demonstrated a trend towards improved diagnostic accuracy over hyperemic MBF.
- Further studies may refine RFR's role in noninvasive FFR assessment.
Background:
Quantitative myocardial perfusion imaging is increasingly used for the diagnosis of coronary artery disease. Quantitative perfusion imaging allows to noninvasively calculate fractional flow reserve (FFR). This so-called relative flow reserve (RFR) is defined as the ratio of hyperemic myocardial blood flow (MBF) in a stenotic area to hyperemic MBF in a normal perfused area. The aim of this study was to assess the value of RFR in the detection of significant coronary artery disease.
Methods And Results:
From a clinical population of patients with suspected coronary artery disease who underwent oxygen-15-labeled water cardiac positron emission tomography and invasive coronary angiography, 92 patients with single- or 2-vessel disease were included. Intermediate lesions (diameter stenosis, 30%-90%; n=75) were interrogated by FFR. Thirty-eight (41%) vessels were deemed hemodynamically significant (>90% stenosis or FFR≤0.80). Hyperemic MBF, coronary flow reserve, and RFR were lower for vessels with a hemodynamically significant lesion (2.01±0.78 versus 2.90±1.16 mL·min(-1)·g(-1); P<0.001, 2.27±1.03 versus 3.10±1.29; P<0.001, and 0.67±0.23 versus 0.93±0.15; P<0.001, respectively). The correlation between RFR and FFR was moderate (r=0.54; P<0.01). Receiver operator characteristic curve analysis showed an area under the curve of 0.82 for RFR, which was not significantly higher compared with that for hyperemic MBF and coronary flow reserve (0.76; P=0.32 and 0.72; P=0.08, respectively).
Conclusions:
Noninvasive estimation of FFR by quantitative perfusion positron emission tomography by calculating RFR is feasible, yet only a trend toward a slight improvement of diagnostic accuracy compared with hyperemic MBF assessment was determined.