Rotational artifacts in on-board cone beam computed tomography
1Department of Medical Physics, The Ottawa Hospital Cancer Centre, Ottawa, Canada.
Abstract:
Rotational artifacts in image guidance systems lead to registration errors that affect non-isocentric treatments and dose to off-axis organs-at-risk. This study investigates a rotational artifact in the images acquired with the on-board cone beam computed tomography system XVI (Elekta, Stockholm, Sweden). The goals of the study are to identify the cause of the artifact, to characterize its dependence on other quantities, and to investigate possible solutions. A 30 cm diameter cylindrical phantom is used to acquire clockwise and counterclockwise scans at five speeds (120 to 360 deg min(-1)) on six Elekta linear accelerators from three generations (MLCi, MLCi2 and Agility). Additional scans are acquired with different pulse widths and focal spot sizes for the same mAs. Image quality is evaluated using a common phantom with an in-house three dimensional contrast transfer function attachment. A robust, operator-independent analysis is developed which quantifies rotational artifacts with 0.02° accuracy and imaging system delays with 3 ms accuracy. Results show that the artifact is caused by mislabelling of the projections with a lagging angle due to various imaging system delays. For the most clinically used scan speed (360 deg min(-1)), the artifact is ∼0.5°, which corresponds to ∼0.25° error per scan direction with the standard Elekta procedure for angle calibration. This leads to a 0.5 mm registration error at 11 cm off-center. The artifact increases linearly with scan speed, indicating that the system delay is independent of scan speed. For the most commonly used pulse width of 40 ms, this delay is 34 ± 1 ms, part of which is half the pulse width. Results are consistent among the three linac generations. A software solution that corrects the angles of individual projections is shown to eliminate the rotational error for all scan speeds and directions. Until such a solution is available from the manufacturer, three clinical solutions are presented, which reduce the rotational error without compromising image quality.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...