Related Experiment Video
Updated: Apr 18, 2026

Evaluating Tests of Cognition using a Computerized Touch-Sensitive Tablet, Eye Tracking, and Functional Magnetic Resonance Imaging
Published on: January 30, 2026
Validating the Accuracy of Reaction Time Assessment on Computer-Based Tablet Devices
Philip Schatz1, Vincent Ybarra2, Donald Leitner2
1Department of Psychology, Saint Joseph's University, Philadelphia, PA, USA schatzSJU@gmail.com.
Tablet timing accuracy varies significantly across devices. Apple iPads and Kindle Fire show minimal delays, while Google Nexus and Samsung tablets exhibit substantial timing errors, impacting reaction time assessments.
Area of Science:
- Cognitive psychology
- Human-computer interaction
- Neuroscience
Background:
- Tablet devices are increasingly used for computer-based assessments.
- Limited research validates the accuracy of tablet-based response timing.
- Understanding timing delays is crucial for reliable data collection.
Purpose of the Study:
- To quantify timing delays between stimulus presentation and touch response on various tablet devices.
- To assess the impact of different operating systems and device models on response timing accuracy.
- To evaluate the implications for clinical reaction time (RT) assessments using tablets.
Main Methods:
- Timing delays were measured on iOS (Apple iPads) and Android (Kindle Fire, Google Nexus, Samsung Galaxy) devices.
- Response intervals tested included 100, 250, 500, and 1,000 milliseconds (ms).
- Simulated touch responses were used to document timing errors.
Main Results:
- Google Nexus and Samsung tablets showed significantly higher timing errors (81-97 ms) compared to Kindle Fire and Apple iPads (27-33 ms).
- iOS 7 demonstrated lower timing error than iOS 6 on Apple devices.
- Simple reaction time trials (250 ms) exhibited 12%-40% error, decreasing to 3%-5% for choice RT trials at 1,000 ms.
Conclusions:
- Significant device-dependent timing discrepancies exist in tablet-based assessments.
- These errors can substantially affect the validity of reaction time measurements, especially in simple RT tasks.
- Calibration of software and hardware is essential for reliable clinical use of tablets in serial assessments.