Related Experiment Video
Updated: Apr 18, 2026

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Adverse tissue reaction to corrosion at the neck-stem junction after modular primary total hip arthroplasty
G Gkagkalis1, P Mettraux2, P Omoumi3
1Departments of Orthopaedics, Centre Hospitalier Universitaire Vaudois, CHUV, University of Lausanne, Switzerland.
Abstract:
Complications related to the neck-stem junction of modular stems used for total hip arthroplasty (THA) are generating increasing concern. A 74-year-old male had increasing pain and a cutaneous reaction around the scar 1 year after THA with a modular neck-stem. Imaging revealed osteolysis of the calcar and a pseudo-tumour adjacent to the neck-stem junction. Serum cobalt levels were elevated. Revision surgery to exchange the stem and liner and to resect the pseudo-tumour was performed. Analysis of the stem by scanning electron microscopy and by energy dispersive X-ray and white light interferometry showed fretting corrosion at the neck-stem junction contrasting with minimal changes at the head-neck junction. Thus, despite dry assembly of the neck and stem on the back table at primary THA, full neck-stem contact was not achieved, and the resulting micromotion at the interface led to fretting corrosion. This case highlights the mechanism of fretting corrosion at the neck-stem interface responsible for adverse local tissue reactions. Clinical and radiological follow-up is mandatory in patients with dual-modular stems.

