Related Experiment Video
Updated: Apr 18, 2026

The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018
A chronoamperometric screen printed carbon biosensor based on alkaline phosphatase inhibition for W(IV) determination
Ana Lorena Alvarado-Gámez1, María Asunción Alonso-Lomillo2, Olga Domínguez-Renedo3
1School of Chemistry & CELEQ, University of Costa Rica, San Pedro de Montes de Oca, 11500-2060 San José, Costa Rica. ana.alvaradogamez@ucr.ac.cr.
Abstract:
This paper presents a chronoamperometric method to determine tungsten in water using screen-printed carbon electrodes modified with gold nanoparticles and cross linked alkaline phosphatase immobilized in the working electrode. Enzymatic activity over 2-phospho-l-ascorbic acid trisodium salt, used as substrate, was affected by tungsten ions, which resulted in a decrease of chronoamperometric current, when a potential of 200 mV was applied on 10 mM of substrate in a Tris HCl buffer pH 8.00 and 0.36 M of KCl. Calibration curves for the electrochemical method validation, give a reproducibility of 5.2% (n = 3), a repeatability of 9.4% (n = 3) and a detection limit of 0.29 ± 0.01 µM. Enriched tap water, purified laboratory water and bottled drinking water, with a certified tungsten reference solution traceable to NIST, gave a recovery of 97.1%, 99.1% and 99.1% respectively (n = 4 in each case) and a dynamic range from 0.6 to 30 µM. This study was performed by means of a Lineweaver-Burk plot, showing a mixed kinetic inhibition.