Related Experiment Video
Updated: Apr 18, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Automatic tracking of gold seed markers from CBCT image projections in lung and prostate radiotherapy
Christopher Gehrke1, Richard Oates1, Prabhakar Ramachandran1
1Peter MacCallum Cancer Centre, Melbourne, Victoria 3002, Australia.
Purpose:
To construct a method and software to track gold seed implants in prostate and lung patients undergoing radiotherapy using CBCT image projections.
Methods:
A mathematical model was developed in the MatLab (Mathworks, Natick, USA) environment which uses a combination of discreet cosine transforms and filtering to enhance several edge detection methods for identifying and tracking gold seed fiducial markers in images obtained from Varian (Varian Medical Systems, Palo Alto, USA) and Elekta (Kungstensgatan, Sweden) CBCT projections.
Results:
Organ motion was captured for 16 prostate patients and 1 lung patient.
Conclusion:
Image enhancement and edge detection is capable of automatically tracking markers for up to 98% (Varian) and 79% (Elekta) of CBCT projections for prostate and lung markers however inclusion of excessive bony anatomy (LT and RT LAT) inhibit the ability of the model to accurate determine marker location.