Staging of keratoconus indices regarding tomography, topography, and biomechanical measurements
Susanne Goebels1, Timo Eppig2, Stefan Wagenpfeil3
1Department of Ophthalmology, Saarland University Medical Center, Homburg/Saar, Germany.
Purpose:
To derive limits of metric keratoconus indices for classification into keratoconus stages.
Design:
Validity and reliability analysis of diagnostic tools.
Methods:
A total of 126 patients from the keratoconus center of Homburg/Saar were evaluated with respect to Amsler criteria, using Pentacam (Keratoconus Index [KI], Topographic Keratoconus Classification [TKC]), Topographic Modeling System (Smolek/Klyce, Klyce/Maeda), and Ocular Response Analyzer (Keratoconus Match Probability [KMP], Keratoconus Match Index [KMI]). Mean value, standard deviation, 90% confidence interval, and the Youden J index for definition of the thresholds were evaluated.
Results:
For separation of keratoconus stages 0/1/2/3/4 we derived the following optimum thresholds: for KI 1.05/1.15/1.31/1.49 and for KMI 0.77/0.32/-0.08/-0.3. For Smolek/Klyce and Klyce/Maeda high standard deviations and overlapping confidence intervals were found; therefore no discrete thresholds could be defined. Nevertheless, for them we still found a good sensitivity and specificity in discriminating between healthy (stage 0) and keratoconus (stages 2-4) eyes in comparison with the other indices.
Conclusions:
We derived thresholds for the metric keratoconus indices KI and KMI, which allow classification of keratoconus stages. These now need to be validated in clinical use. Smolek/Klyce and Klyce/Maeda were not sufficiently sensitive to allow classification into individual stages, but these indices did show a good specificity and sensitivity in discriminating between keratoconus and healthy eyes.