Metal and motion artifacts by cone beam computed tomography (CBCT) in dental and maxillofacial study
Cosimo Nardi1, Claudia Borri, Francesco Regini
1Department of Experimental and Clinical Biomedical Sciences Radiodiagnostic Unit n. 2, University of Florence-Azienda Ospedaliero-Universitaria Careggi, Largo Brambilla 3, 50134, Florence, Italy, cosimo.nardi@unifi.it.
Objectives:
This study aimed at evaluating incidence/degree of metal/motion artifacts and CT-dose-index in oral/maxillofacial examinations using Cone-Beam-CT.
Methods:
Interferences caused by metal and motion artifacts were evaluated in 500 patients aged from 6 to 81 years, in dental arches, maxillofacial and splanchocranium Cone-Beam-CT exams. The interferences was divided into four progressive degrees (G0-G3) related to the possibility to answer the clinical query. The parameters considered were field-of-view, scan time, patient's age, and anatomical area. Furthermore volumetric CT-dose-index was measured.
Results:
In the presence of metal artifacts the clinical query was always answered (G3 = 0). No artifacts (G0) were found in all cases when metal was beyond 5 cm from interest site and in 18.4% when metal was inside this distance. Beam hardening and photon starvation due to implants, restoration and orthodontic therapies achieved 56.2% G1 and 25.4% G2. Motion artifacts were more frequent in under ten (31.5%) and over sixty (82.2%), and in mandible analysis (inferior arch 59.5%, both arches 47.3%). Moreover, their incidence and intensity were influenced by scan time (49.1% at 36 s) but not by field-of-view. Mean volumetric CT-dose-index of all patients was mGy 9.11 (mGy 3.62, 5.78, 8.89, and 13.07 at 18, 24, 26, and 36 s, respectively).
Conclusions:
In our series Cone-Beam-CT diagnostic evaluation was never inhibited by metal artifacts and only in 1.9% of the cases by motion artifacts, always with a very low CT-dose-index.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...