A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Accuracy of measurement of hand compartment pressures: a cadaveric study
Justin C Wong1, Michael M Vosbikian1, Joseph M Dwyer1
1Department of Orthopaedic Surgery, Thomas Jefferson University Hospital, Philadelphia, PA; Department of Orthopaedic Surgery, The Rothman Institute at Thomas Jefferson University Hospital, Philadelphia, PA.
Purpose:
To determine the accuracy of digital palpation for clinical assessment of elevated intracompartmental pressure compared with needle manometry in a simulated compartment syndrome of the hand.
Methods:
Three cadaveric hands were configured with interstitial fluid infusion and an arterial line pressure monitor to create and continuously measure intracompartmental pressure in the thenar and hypothenar compartments. Seventeen assessors clinically judged the presence or absence of compartment syndrome based on digital palpation for firmness and then measured pressures with a handheld manometer. An intracompartmental pressure threshold of 30 mm Hg or greater was used to diagnose compartment syndrome.
Results:
The sensitivity and specificity of digital palpation of the thenar eminence were 49% and 79%, respectively, with a positive predictive value (PPV) of 86% and negative predictive value (NPV) of 37%. Using the handheld manometer, the sensitivity and specificity increased to 97% and 86% with a PPV of 95% and NPV of 92%. The sensitivity and specificity of digital palpation of the hypothenar eminence were 62% and 83%, respectively, with improvement of 100% and 100%, respectively, with a handheld manometer. For the hypothenar compartment, use of a handheld manometer improved the PPV from 92% to 100% and the NPV from 40% to 100% compared with digital palpation.
Conclusions:
Digital palpation alone was insufficient to detect elevated compartment pressures in hands at risk for compartment syndrome. Handheld invasive pressure measurement was a useful adjunct for detecting elevated interstitial tissue pressures and may aid in diagnosing compartment syndrome.
Type Of Study/Level Of Evidence:
Diagnostic II.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...