Antimicrobial Release From Prefabricated Spacers Is Variable and the Dose Is Low
Oren Goltzer1, Alex McLaren, Derek Overstreet
1Banner Good Samaritan Medical Center, Banner Orthopaedic Residency, 1320 North 10th Street, Suite A, Phoenix, AZ, 85006, USA.
Clinical Orthopaedics and Related Research
|February 5, 2015
Summary
Antimicrobial release from prefabricated bone cement spacers is comparable to hand-mixed formulations. Release is consistent across surfaces and components, supporting their use in treating bone infections.
Area of Science:
- Orthopaedic Surgery
- Biomaterials Science
- Infectious Disease Management
Background:
- High-dose antimicrobial-loaded bone cement (ALBC) is standard for orthopaedic infections.
- Elution characteristics of prefabricated ALBC spacers are not well understood compared to hand-mixed ALBC.
Purpose of the Study:
- Compare antimicrobial release over time between prefabricated and hand-mixed ALBC.
- Assess uniformity of release across prefabricated spacer surfaces.
- Investigate variations between different prefabricated spacer components.
- Determine if textured surfaces yield different release rates.
Main Methods:
- Comparative laboratory study of six prefabricated ALBC spacer components (hip and knee) and three hand-mixed ALBC formulations.
- Gentamicin elution measured from 41 sites on prefabricated spacers and 15 specimens of hand-mixed ALBC.
- Statistical analysis using ANOVA and Tukey's multiple-comparison test or t-test.
Main Results:
- Both prefabricated and hand-mixed ALBC showed highest gentamicin release in the first 24 hours, decreasing over 7 days.
- At Day 7, prefabricated spacers eluted more than 1g hand-mixed ALBC but less than 5g or 10g formulations.
- Release varied across spacer surfaces (2.7-9.9 µg/mm²) and between components (3.5-5.5 µg/mm²), with no significant differences found.
- Release from textured surfaces was inconsistent.
Conclusions:
- Antimicrobial release from prefabricated ALBC spacers is consistent with low-dose ALBC formulations.
- Variations in release across surfaces and between components are minimal relative to antimicrobial load.
- Prefabricated spacers offer consistent local antimicrobial delivery, similar to other low-dose ALBC methods.
Related Concept Videos
Bacterial Signaling
43.7K
Bacterial signaling can occur within bacteria (intracellular) or between bacteria (intercellular). At times, a group of bacteria behaves like a community. To achieve this, they engage in quorum sensing, the perception of higher cell density that causes changes in gene expression. Quorum sensing involves both extracellular and intracellular signaling. The signaling cascade starts with a molecule called an autoinducer (AI). Individual bacteria produce AIs that move out of the bacterial cell...
43.7K
Antimicrobial Effectiveness
2.1K
The effectiveness of antimicrobial agents depends on various factors influencing their ability to eliminate microbial populations. Larger microbial populations require more time for complete eradication, emphasizing the importance of population size analysis when evaluating antimicrobial efficacy.Microbial resistance to antimicrobial agents varies significantly. Highly resilient microorganisms include endospores, gram-negative bacteria, and non-enveloped viruses, while prions are exceptionally...