Heart-rate dependent improvement in image quality and diagnostic accuracy of coronary computed tomographic
Iksung Cho1, Kimberly Elmore2, Bríain Ó Hartaigh2
1Department of Radiology, NewYork-Presbyterian Hospital and the Weill Cornell Medical College, New York, NY, USA; Division of Cardiology, Severance Cardiovascular Hospital and Severance Biomedical Science Institute, Yonsei University College of Medicine, Yonsei University Health System, Seoul, South Korea.
Background:
To determine the effect of a novel intracycle motion correction algorithm (MCA) on diagnostic accuracy of coronary computed tomographic angiography.
Methods:
Coronary artery phantom models were scanned at static and heart rate (HR) simulation of 60-100 beat/min and reconstructed with a conventional algorithm and MCA.
Results:
Among 144 coronary segments, improvements in image interpretability, quality, and diagnostic accuracy by MCA were observed for HRs of 80 and 100 (P<.05 for all), but not for HR of 60.
Conclusion:
Novel intracycle MCA demonstrates improved HR-dependent image interpretability, and quality and accuracy, particularly at higher HRs.