Accelerated acquisition of tagged MRI for cardiac motion correction in simultaneous PET-MR: phantom and patient

Chuan Huang1, Yoann Petibon2, Jinsong Ouyang3

  • 1Center for Advanced Medical Imaging Sciences, Division of Nuclear Medicine and Molecular Imaging, Department of Imaging, Massachusetts General Hospital, Boston, Massachusetts 02114; Department of Radiology, Harvard Medical School, Boston, Massachusetts 02115; and Departments of Radiology, Psychiatry, Stony Brook Medicine, Stony Brook, New York 11794.

Medical Physics
|February 6, 2015
PubMed
Abstract

Insights

Accelerated tagged MR (tMR) acquisition significantly reduces scan times for cardiac PET-MR motion correction. This method maintains diagnostic image quality, making cardiac PET-MR more feasible for routine clinical use.

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Cardiovascular Imaging

Background:

  • Cardiac and respiratory motion degrade PET image quality, impacting diagnostic accuracy.
  • Tagged MR (tMR)-based PET motion correction improves myocardial defect detection but requires long acquisition times.

Purpose of the Study:

  • To develop and evaluate an accelerated tMR acquisition strategy for PET motion correction.
  • To assess the impact of accelerated tMR on motion-corrected PET image quality using phantom and patient data.

Main Methods:

  • Retrospective acceleration of tMR data using parallel imaging (GRAPPA) and compressed sensing (kt-FOCUSS).
  • Motion field estimation via nonrigid B-spline registration on accelerated and fully sampled tMR images.
  • Incorporation of motion fields into motion-corrected ordered subset expectation maximization (OSEM) reconstruction with motion-dependent attenuation correction.

Main Results:

  • Accelerated tMR introduced artifacts but yielded comparable PET image quality to fully sampled tMR at acceleration factors <4.
  • Quantitative analysis showed similar myocardial defect contrast.
  • Even at high acceleration factors, motion-corrected PET images were superior to uncorrected images.

Conclusions:

  • Accelerated tMR (4x acceleration) provides accurate motion fields for PET motion correction.
  • This acceleration reduces tMR acquisition time, enhancing compatibility with routine clinical cardiac PET-MR.
  • The method preserves diagnostic image quality for clinical applications.