Related Experiment Video
Updated: Apr 17, 2026

Dynamic Navigation for Dental Implant Placement
Published on: September 13, 2022
Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator
Sreeha Kaluva Kolanu1, Nagesh Bolla2, Sujana Varri3
1Post Graduate, Master of Dental Surgery in Speciality of Conservative Dentistry & Endodontics, Sibar Institute of Dental Sciences , Guntur, Andhra Pradesh, India .
Aim:
AIM of this study is to evaluate the influence of critical diameter of apical foramen and file size using propex pixi apex locator in working length determination.
Materials And Methods:
In this study, ten single rooted teeth were selected. They were decoronated at cemento enamel junction. After determining the actual working length, they were embedded in alginate mold. Foramina were widened from 0.6mm to 0.8mm. The measurements were taken with electronic apex locator propex pixi with files from sizes 10 K to respective sizes. Statistical accuracy of propex pixi was calculated by using Anova test for different diameters and for the influence of file size.
Results:
RESULTS showed that propex pixi apex locator was accurate when foramen diameter is 0.6 (60k file size), its accuracy diminished with increased foramen diameter
Conclusion:
Propex pixi is accurate for foramen diameter of 0.6mm, independent of file size. Its accuracy decreases as apical foramen widens, so care should be taken when using clinically.