Editor's choice--Angulation of the C-arm during complex endovascular aortic procedures increases radiation exposure
M A Albayati1, S Kelly1, D Gallagher2
1Academic Department of Vascular Surgery, Cardiovascular Division, King's College London, BHF Centre of Research Excellence & NIHR Biomedical Research Centre at King's Health Partners, St Thomas' Hospital, London, UK.
Objectives/Background:
The increased complexity of endovascular aortic repair necessitates longer procedural time and higher radiation exposure to the operator, particularly to exposed body parts. The aims were to measure directly exposure to radiation of the bodies and heads of the operating team during endovascular repair of thoracoabdominal aortic aneurysms (TAAA), and to identify factors that may increase exposure.
Methods:
This was a single-centre prospective study. Between October 2013 and July 2014, consecutive elective branched and fenestrated TAAA repairs performed in a hybrid operating room were studied. Electronic dosimeters were used to measure directly radiation exposure to the primary (PO) and assistant (AO) operator in three different areas (under-lead, over-lead, and head). Fluoroscopy and digital subtraction angiography (DSA) acquisition times, C-arm angulation, and PO/AO height were recorded.
Results:
Seventeen cases were analysed (Crawford II-IV), with a median operating time of 280 minutes (interquartile range 200-330 minutes). Median age was 76 years (range 71-81 years); median body mass index was 28 kg/m(2) (25-32 kg/m(2)). Stent-grafts incorporated branches only, fenestrations only, or a mixture of branches and fenestrations. A total of 21 branches and 38 fenestrations were cannulated and stented. Head dose was significantly higher in the PO compared with the AO (median 54 μSv [range 24-130 μSv] vs. 15 μSv [range 7-43 μSv], respectively; p = .022), as was over-lead body dose (median 80 μSv [range 37-163 μSv] vs. 32 μSv [range 6-48 μSv], respectively; p = .003). Corresponding under-lead doses were similar between operators (median 4 μSv [range 1-17 μSv] vs. 1 μSv [range 1-3 μSv], respectively; p = .222). Primary operator height, DSA acquisition time in left anterior oblique (LAO) position, and degrees of LAO angulation were independent predictors of PO head dose (p < .05).
Conclusions:
The head is an unprotected area receiving a significant radiation dose during complex endovascular aortic repair. The deleterious effects of exposure to this area are not fully understood. Vascular interventionalists should be cognisant of head exposure increasing with C-arm angulation, and limit this manoeuvre.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging