Related Experiment Video
Updated: Apr 17, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Dislodgement of coronary stent due to rupture of stent balloon
Burak Ayça1, Ertuğrul Okuyan2, İrfan Şahin2
1Department of Cardiology, Bağcılar Training and Research Hospital, İstanbul, Turkey. drburakayca@yahoo.com.tr.
Abstract:
Rare stent complications, including dislodgement of stent, unexpanded stent, stent fracture and stent loss etc. can occur during percutaneous coronary interventions (PCI). We present a semi-expanded and dislodged stent due to rupture of stent balloon during primary PCI in this case report. An interventional cardiologist should be aware of the possibility of rare complications, such as in this case, and have enough experience and knowledge to handle them.
Related Concept Videos
Aneurysm III: Interprofessional Care
Coronary Artery Disease II: Pathophysiology
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Coronary Artery Disease V: Interprofessional Care
Aneurysm I: Introduction
Acute Coronary Syndrome I: Introduction