A fast method for measuring psychophysical thresholds across the cochlear implant array.
Julie A Bierer1, Steven M Bierer2, Heather A Kreft3
1Department of Speech and Hearing Sciences, University of Washington, Seattle, WA, USA jbierer@u.washington.edu.
Trends in Hearing
|February 7, 2015
Summary
A new rapid threshold measurement procedure for cochlear implants (CIs) offers a faster way to estimate hearing thresholds. While slightly less reliable than traditional methods, its speed makes it promising for clinical use.
Area of Science:
- Audiology
- Neuroscience
- Biomedical Engineering
Background:
- Cochlear implants (CIs) require accurate threshold measurements for optimal hearing rehabilitation.
- Traditional threshold measurement procedures can be time-consuming.
- Developing rapid and reliable methods is crucial for efficient clinical practice.
Purpose of the Study:
- To propose and evaluate a novel, rapid threshold measurement procedure for cochlear implant users.
- To compare the proposed Bekesy tracking-based method with a traditional adaptive procedure.
- To assess the speed and reliability of the new method.
Main Methods:
- A Bekesy tracking procedure with virtual spectral sweeps using current steering was developed.
- Absolute thresholds for biphasic pulse trains were measured in 15 postlingually deafened adult CI users.
- Thresholds were compared between the new tracking procedure and a traditional forced-choice adaptive procedure under monopolar and quadrupolar stimulation.
Main Results:
- No systematic differences were observed between threshold estimates from the new sweep procedure and the traditional adaptive procedure.
- The new method was approximately four times faster than the traditional procedure for the same number of repetitions.
- Test-retest reliability of the new sweep procedure was slightly poorer compared to the traditional procedure.
Conclusions:
- The proposed rapid threshold measurement procedure demonstrates potential for clinical application due to its significant speed advantage.
- The trade-off between speed and reliability suggests careful consideration for specific clinical scenarios.
- Faster threshold estimation is particularly valuable when combined with focused stimulation techniques in cochlear implant fitting.