Bioreactance is a reliable method for estimating cardiac output at rest and during exercise
T W Jones1, D Houghton2, S Cassidy2
1Institute of Neurosciences and Newcastle University Institute for Ageing, Newcastle University, Newcastle upon Tyne NE2 4HH, UK.
Background:
Bioreactance is a novel noninvasive method for cardiac output measurement that involves analysis of blood flow-dependent changes in phase shifts of electrical currents applied across the thorax. The present study evaluated the test-retest reliability of bioreactance for assessing haemodynamic variables at rest and during exercise.
Methods:
22 healthy subjects (26 (4) yrs) performed an incremental cycle ergometer exercise protocol relative to their individual power output at maximal O2 consumption (Wmax) on two separate occasions (trials 1 and 2). Participants cycled for five 3 min stages at 20, 40, 60, 80 and 90% Wmax. Haemodynamic and cardiorespiratory variables were assessed at rest and continuously during the exercise protocol.
Results:
Cardiac output was not significantly different between trials at rest (P=0.948), or between trials at any stage of the exercise protocol (all P>0.30). There was a strong relationship between cardiac output estimates between the trials (ICC=0.95, P<0.001) and oxygen consumption (ICC=0.99, P<0.001). Stroke volume was also not significantly different between trials at rest (P=0.989) or during exercise (all P>0.15), and strong relationships between trials were found (ICC=0.83, P<0.001).
Conclusions:
The bioreactance method demonstrates good test-retest reliability for estimating cardiac output at rest and during different stages of graded exercise testing including maximal exertion.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...