Related Experiment Video
Updated: Apr 17, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Comparison of visual outcomes of 2 diffractive trifocal intraocular lenses
Eduardo F Marques1, Tiago B Ferreira1
1From the Cruz Vermelha Hospital (Marques) and Egas Moniz Hospital (Ferreira), Lisbon, Portugal.
Purpose:
To compare the visual outcomes after cataract surgery with bilateral implantation of 1 of 2 diffractive trifocal intraocular lenses (IOLs).
Setting:
Two clinical centers, Lisbon, Portugal.
Design:
Prospective comparative case series.
Methods:
Phacoemulsification with bilateral implantation of a Finevision Micro F IOL (Group 1) or an AT Lisa tri 839 MP IOL (Group 2) was performed. Over a 3-month follow-up, the main outcome measures were uncorrected distance visual acuity (UDVA), corrected monocular and binocular distance visual acuity, uncorrected intermediate visual acuity at 80 cm, distance-corrected intermediate visual acuity (DCIVA), uncorrected near visual acuity at 40 cm, distance-corrected near visual acuity (DCNVA), spherical equivalent (SE) refraction, defocus curves, contrast sensitivity, presence of dysphotopsia, and use of spectacles.
Results:
Each group comprised 30 eyes (15 patients). The mean values at 3 months were UDVA, 0.03 logMAR ± 0.08 (SD) (Group 1) and 0.08 ± 0.12 (Group 2) (P = .765); DCIVA, 0.04 ± 0.07 logMAR and 0.18 ± 0.18 logMAR, respectively (P = .048); DCNVA, 0.03 ± 0.06 logMAR and 0.11 ± 0.08 logMAR, respectively (P = .032); SE, -0.25 ± 0.30 diopter (D) and -0.02 ± 0.39 D, respectively (P = .087). There was no significant difference in contrast sensitivity or dysphotopic phenomena between groups.
Conclusions:
Both trifocal IOL models provided excellent distance, intermediate, and near visual outcomes. Monocular DCIVA and DCNVA appeared slightly better in Group 1. Predictability of the refractive results and optical performance were excellent, and all patients achieved spectacle independence.
Financial Disclosure:
Neither author has a financial or proprietary interest in any material or method mentioned.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...