Protocol for Measuring the Thermal Properties of a Supercooled Synthetic Sand-water-gas-methane Hydrate Sample
Published on: March 21, 2016
Temperature manifold for a stopped-flow machine to allow measurements from -10 to +40°C
Jonathan Walklate1, Michael A Geeves1
1School of Biosciences, University of Kent, Canterbury, Kent CT2 7NJ, UK.
Abstract:
Conducting enzymatic stopped-flow experiments at temperatures far removed from ambient can be very problematic because extremes in temperature (<10°C or >30°C) can damage the machine or the enzyme. We have devised a simple manifold that can be attached to most commercial stopped-flow systems that is independently heated or cooled separate from the main stopped-flow system. Careful calibration of the flow circuit allows the sample to be heated or cooled to the measurement temperature (-8 to +40°C) 1 to 2s before mixing in the reaction chamber. This approach allows measurements at temperatures where the stopped flow or the protein is normally unstable. To validate the manifold, we investigated the well-defined ATP-induced dissociation of rabbit muscle myosin subfragment 1 (S1) from its complex with pyrene-labeled actin. This process has both temperature-dependent and -independent components. Use of ethylene glycol allowed us to measure the reaction below 0°C and up to 42°C, and as expected the second-order rate constant (K1k+2) and the maximum rate of dissociation (k+2) both increased with temperature, whereas 1/K1 is unaffected by the change in temperature.
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Thermometers and Temperature Scales
As many physical properties depend on temperature, the variety of thermometers is...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...