Related Experiment Video
Updated: Apr 17, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Comparison between Glidescope and Lightwand for tracheal intubation in patients with a simulated difficult airway
Ki-Hwan Yang1, Chan Ho Jeong1, Kyung Chul Song1
1Department of Anesthesiology and Pain Medicine, Inha University School of Medicine, Incheon, Korea.
Background:
Although Lightwand and Glidescope have both shown high success rates for intubation, there has been no confirmation as to which device is most effective for difficult endotracheal intubation. We compared the Glidescope and Lightwand devices in terms of duration of intubation and success rate at the first attempt in a simulated difficult airway situation.
Methods:
Fifty-eight patients were randomized to undergo tracheal intubation with either the Glidescope (Glidescope group, n = 29) or the Lightwand (Lightwand group, n = 29). All patients were fitted with a semi-hard cervical collar in order to simulate a difficult airway, and intubation was attempted with the assigned airway device. The data collected included the rate of successful endotracheal intubation, the number of attempts required, the duration of the intubation, as well as the interincisor distance, hemodynamic variables, and adverse effects.
Results:
There was no difference between Glidescope group (92.6%) and Lightwand group (96.4%) in terms of success rate for the first attempt at intubation. The duration of successful intubation for the first tracheal intubation attempt was significantly longer in Glidescope group than in Lightwand group (46.9 sec vs 29.5 sec, P = 0.001). All intubations were completed successfully within two intubation attempts. The incidence of hypertension was significantly higher in Glidescope group than in Lightwand group (51.9% vs 17.9%, P = 0.008).
Conclusions:
In a simulated difficult airway situation, endotracheal intubation using Lightwand yielded a shorter duration of intubation and lower incidence of hypertension than when using Glidescope.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Suctioning II: Procedure