Purpose:
To compare the visual and optical outcomes of four multifocal intraocular lenses (IOLs) with three different near additions of +3.00 diopters (D), +3.75 D and +4.00 D.
Methods:
In this prospective study, 133 eyes of 88 patients were implanted with one of the following IOLs: AcrySof(®) ReSTOR(®) SN6AD1 (+3.00 D) for Group A, Acriva(UD) Reviol BB MF 613 or BB MFM 611 (+3.75 D) for Group B, and AcrySof(®) ReSTOR(®) SN6AD3 (+4.00 D) for Group C. The visual acuity, refraction, intraocular pressure, tomography and corneal endothelial cell density (ECD) were compared between the three groups preoperatively and at 6 month postoperatively. Defocus curve, contrast sensitivity and higher order aberrations (HOAs) at 6 month postoperative visit were measured and compared.
Results:
There were no statistically significant differences in distance visual acuity, refraction, intraocular pressure or ECD among the three groups after 6 months (P > 0.05). The photopic contrast sensitivity in Group C was statistically better than in Group A (P < 0.05). The scotopic ocular aberration in Group B was statistically greater compared to that in Group A (P < 0.05). The highest near-visual peaks were -0.06 logMAR at a -2.50 D (40 cm) in Group A, -0.07 logMAR at -3.00D (33 cm) in Group B, and -0.06 logMAR at -3.50 D (29 cm) in Group C. Statistically significant differences in near and intermediate visual acuities were observed among the three groups at -2.00 D (50 cm), -2.50 D (40 cm), -3.50 D (29 cm) and -4.00 D (25 cm) (P < 0.01).
Conclusion:
AcrySof(®) ReSTOR(®) SN6AD1 IOLs (+3.00 D) and SN6AD3 (+4.00 D) IOLs provided the best intermediate and near vision, respectively. Both intermediate and near vision were comparatively better in the eyes with Acriva(UD) Reviol BB MFM 611 IOLs or BB MF 613 IOLs (+3.75 D).
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Depth Perception and Spatial Vision
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...