A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Comparison of parenchymal and ventricular intracranial pressure readings utilizing a novel multi-parameter
Tracey Berlin1, Cristina Murray-Krezan2, Howard Yonas3
1Department of Neurosurgery, University of New Mexico Hospital, MSC10 5615, 1 University of New Mexico, Albuquerque, NM 87131-0001 USA.
Introduction:
Both ventricular and parenchymal devices are available for measurement of intracranial pressure (ICP). The Hummingbird(®) Synergy Ventricular System is a novel device allowing multi-parametric neurological monitoring, including both ventricular and parenchymal ICP. The purpose of this study is to compare the congruence of the device's ventricular and parenchymal ICP readings.
Methods:
This single-center, quantitative, interventional study compared parenchymal and ventricular ICP readings from 35 patients with the Hummingbird(®) System. If a difference of > ± 3 mmHg existed between an individual patient's parenchymal and ventricular values, progressive intervention strategies were applied to correct identified issues.
Results:
From a total of 2,259 observations, statistical analysis revealed congruence (within ±0-3 mmHg) of 93% of readings comparing parenchymal and ventricular ICP. Of the observations requiring intervention, 58% involved the parenchymal component, 30% involved the ventricular component, and 12% involved both components. Following prescribed interventions, 98% of readings became congruent (within ±0-3 mmHg). The adjusted mean difference between the two methods was -0.95 (95% CI: -0.97,-0.93) mmHg and all mean ICP readings fell between -2 and 2 mmHg.
Conclusion:
The Hummingbird(®) Synergy Ventricular System demonstrates congruence between ventricular and parenchymal ICP measurements within accepted parameters. Interventions required to realign parenchymal and ventricular readings serve as reminders to clinicians to be vigilant with catheter/cable connections and to maintain appropriate positioning of the ventricular drainage system. The results of this study support the recommendation to use the parenchymal ICP component for routine ICP monitoring, allowing dedication of the ventricular catheter to drainage of cerebrospinal fluid (CSF).
Related Concept Videos
Increased Intracranial Pressure l: Introduction
Increased Intracranial Pressure ll: Pathophysiology
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement