Optimization of phase-contrast MRI for the quantification of whole-brain cerebral blood flow
Shin-Lei Peng1,2,3, Pan Su1,2,4, Fu-Nien Wang3
1Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
Background:
Whole-brain cerebral blood flow (CBF) measured by phase-contrast MRI (PC-MRI) provides an important index for brain function. This work aimed to optimize the PC-MRI imaging protocol for accurate CBF measurements.
Methods:
Two studies were performed on a 3 Tesla system. In Study 1 (N = 12), we optimized in-plane resolution of PC-MRI acquisition for CBF quantification by considering accuracy, precision, and scan duration. In Study 2 (N = 7), we assessed the detrimental effect of nonperpendicular imaging slice orientation on CBF quantification. Both One-way analysis of variance with repeated measurement and Friedman test were used to examine the effects of resolution and angulation on CBF quantification. Additionally, we evaluated the inter-rater reliability in PC-MRI data processing.
Results:
Our results showed that CBF measurement with 0.7 mm resolution could be overestimated by up to 13.3% when compared with 0.4 mm resolution. Moreover, CBF could also be overestimated by up to 18.8% when the slice orientation is deviated by 30° from the ideal angulation. However, within 10° of the ideal slice orientation, estimated CBF was not significantly different from each other (P = 0.23 and 0.45 for internal carotid artery and vertebral artery, respectively). Inter-rater difference was <3%.
Conclusion:
For fast and accurate quantification of whole-brain CBF with PC-MRI, we recommend the use of an imaging resolution of 0.5 mm and a slice orientation that is less than 10° from vessel's axial plane.