Bone conduction noise exposure via ventilators in the neonatal intensive care unit
Mohammad Abraham Kazemizadeh Gol1,2, Angela Black1,2,3, James Sidman1,2
1Children's Hospitals and Clinics of Minnesota, Minneapolis, Minnesota.
Objectives/Hypothesis:
To demonstrate that neonatal ventilators can expose patients to high noise levels through bone conduction (BC) as well as air conduction (AC).
Study Design:
Observational study.
Methods:
Three ventilators and various settings on a positive airway pressure machine (continuous, high bilevel, and low bilevel pressure) were tested. A sound level meter was used to measure the noise levels at a set distance from the ventilator to represent AC, on the ventilator circuit to represent BC at the alveolus, and within the ventilator circuit.
Results:
The BC sound levels (74.1, 81.1, 86, 89.2 dBC) were significantly higher than the AC sound levels (72.8, 72.9, 70, 71.7 dBC) for the jet ventilator, continuous positive airway pressure setting, low bilevel setting, and high bilevel setting, respectively (P < .001). The sound level within the ventilator circuit ranged from 94.9 to 113.2 dBC depending on the machine/setting and was significantly louder than both AC or BC for all machines/settings (P < .001).
Conclusions:
There are concerning ventilator dependent noise levels present on and within ventilation circuitry that could be presented to the infant via BC.
Level Of Evidence:
NA
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...