Related Experiment Video
Updated: Apr 17, 2026

Evaluation of Patients' Posture and Gait Profile After Lumbar Fusion Surgery by Video Rasterstereography and Treadmill Gait Analysis
Published on: March 23, 2019
Quantitative, Comparative Assessment of Gait Between Single-Radius and Multi-Radius Total Knee Arthroplasty Designs
Bethany Larsen1, Marc C Jacofsky1, David J Jacofsky1
1The CORE Institute, Phoenix, Arizona.
Abstract:
Gait of single-radius (SR, n=16) and multi-radius (MR, n=16) posterior stabilized total knee arthroplasties was compared, along with controls (n=16), pre-op and 1 year post-op. Computer navigation and standard order sets controlled confounding variables. Post-operatively, SR knees did not differ from controls while MR knees continued to differ in important knee kinetic and kinematic properties. MR knees remained more extended (P=0.019) and had decreased power absorption (P=0.0001) during weight acceptance compared to the SR knees. Both surgical groups had similar KSS for Knee Scores (P=0.22) and Function Scores (P=0.58). The significant biomechanical differences are likely influenced by patella-femoral moment arm geometry and changing ligament laxity throughout the active range of motion.