Related Experiment Video
Updated: Apr 17, 2026

08:30
X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
15.0K
Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study
Akmal Sabarudin1, Zakira Mustafa, Khadijah Mohd Nassir
1Diagnostic Imaging & Radiotherapy Program School of Diagnostic & Applied Health Sciences Faculty of Health Sciences Universiti Kebangsaan Malaysia 50300 Kuala Lumpur, Malaysia.. akmal.sabarudin@ukm.edu.my.
Journal of Applied Clinical Medical Physics
|February 14, 2015
Summary
Tube current modulation (TCM) significantly reduces radiation dose in thoracic and abdomen-pelvic CT scans. This technique is highly recommended for routine use to lower patient exposure during these common imaging procedures.
Area of Science:
- Medical Imaging
- Radiological Physics
- Radiation Dosimetry
Background:
- Computed Tomography (CT) scans deliver significant radiation doses.
- Optimizing radiation dose while maintaining image quality is crucial in CT imaging.
- Tube current modulation (TCM) is a technique used to adjust radiation output during CT scans.
Purpose of the Study:
- To compare radiation doses in thoracic and abdomen-pelvic CT scans.
- To evaluate the effectiveness of tube current modulation (TCM) in reducing radiation dose.
- To assess dose reduction using effective dose (ED) and size-specific dose estimation (SSDE).
Main Methods:
- A phantom study was conducted to simulate thoracic and abdomen-pelvic CT scans.
- Radiation doses were measured with and without the use of tube current modulation (TCM).
- Absorbed doses to radiosensitive organs were measured using thermoluminescence dosimeter-100 (TLD-100).
- Effective dose (ED) and size-specific dose estimation (SSDE) were calculated.
Main Results:
- TCM significantly reduced both ED and SSDE in thoracic and abdomen-pelvic CT scans compared to protocols without TCM.
- With TCM, ED for thoracic CT was 6.50 ± 0.29 mSv and for abdomen-pelvic CT was 6.01 ± 0.20 mSv.
- Without TCM, ED increased to 20.07 ± 0.24 mSv (thoracic) and 17.30 ± 0.41 mSv (abdomen-pelvic).
- Highest absorbed doses were observed in the breast (thoracic CT) and skin (abdomen-pelvic CT), with substantial reductions when TCM was used.
Conclusions:
- Tube current modulation (TCM) is highly effective in reducing radiation dose during thoracic and abdomen-pelvic CT.
- The use of TCM leads to significant dose savings, particularly for radiosensitive organs.
- TCM is strongly recommended for routine implementation in clinical practice for thoracic and abdomen-pelvic CT examinations.