Abstract:
Application of half bridge topology for ultrasonic transducer excitation using long pulse trains is presented. The novelty of the approach is the high speed solution for a high side drive. A commercially available high speed digital isolator and a high speed MOSFET driver were combined to give the possibility to deliver fast driving signals to a high side N-channel MOSFET. The experimental investigation indicates that the output amplitude of the fundamental harmonic can reach 624 Vp-p for light loads and 552 Vp-p when driving 50 Ω loads. The operation frequency at such voltages can reach 10 MHz for unloaded or 50 Ω load condition and 6 MHz when driving capacitive 3000 pF loads. The output impedance is 13 Ω for voltages below 500 Vp-p and 16-26 Ω for voltages 500 Vp-p and above.