Related Experiment Video
Updated: Apr 17, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Elastic limits in torsion of reciprocating nickel-titanium instruments
Jung-Hong Ha1, Seo-Ryeong Kim2, Antheunis Versluis3
1Department of Conservative Dentistry, School of Dentistry, Kyungpook National University, Daegu, Korea.
Introduction:
Reciprocating angle and torsional load at the superelastic limit were investigated for nickel-titanium (NiTi) instruments used with reciprocating movements.
Methods:
Two reciprocating NiTi instruments (Reciproc R25 [VDW, Munich, Germany] and WaveOne Primary [Dentsply Maillefer, Ballaigues, Switzerland]) and ProTaper F2 (Dentsply Maillefer) were tested for 2 torsional conditions using a custom-designed testing device. Rotational angles were applied, and generated torque values were recorded. The first test condition fixed the files at 5 mm and repetitively rotated them with gradually increasing angles up to 250° (n = 10). The second test subjected the files to a single continuous rotation until fracture (n = 10). The superelastic limits of the instruments were determined from their torque-rotation curves. Statistical analysis was performed (Kruskal-Wallis) at a 95% significance level. The tested specimens were examined under a scanning electron microscope.
Results:
It was found that the angle at the superelastic limit was higher for R25 and Primary files than the ProTaper file (P < .05). Repetitive torsional loading with a gradually increasing rotational angle reduced the torsional resistance compared with the single rotation motion. At the 5-mm fixation level all files had superelastic limit angles higher than the 170° set in the dedicated reciprocating motor. The scanning electron microscopic analysis showed features of torsional failure.
Conclusions:
Under the conditions of this study, the 170° set angle of dedicated motors for reciprocating file systems is safe at the 5-mm level. Reuse of reciprocating systems, even with the movements within the elastic limit, may cause deterioration of the instruments.
More Related Videos
Related Concept Videos
Circular Shafts - Elastoplastic Materials
As torque on the...
Stress Concentrations in Circular Shafts
Angle of Twist - Elastic Range
Torsion of Noncircular Members
Circular Shaft - Stresses in Linear Range
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as...