A new approach for decreasing the detection limit for a ketamine(I) ion-selective electrode
Hazem M Abu Shawish1, Hassan Tamous2, Salman M Saadeh3
1Chemistry Department, College of Sciences, Al-Aqsa University, Gaza, Palestine.
Abstract:
Our endeavors of lowering the detection limit for a ketamine(I) ion-selective electrode were described. The paper stresses the electrode which showed best results for determination of ketamine ion. The present electrode incorporates ketamine-phosphomolybdate (KT-PM) as ion-exchanger combined with the lipophilic anionic additive (Na-TPB) dissolved in dibutyl phthalate (DBP) as a plasticizer. The characteristics of the electrode were elaborately measured and its performance was tested in various samples and urine. It has favorable features as it provides measurements of the potential with a near-Nernstian slope of 56.6±0.3mV/decade over the concentration range of 1.5×10(-6)-1.0×10(-2)M over the pH range 3.0-6.8 in a short response time (7s). Importantly, it has a low detection limit of 1.2×10(-7)M and its life-span is 22days. Moreover, it displayed notable selectivity for ketamine ion over other species such as inorganic and organic cations and different excipients which may be present in pharmaceutical preparations. The sensor was applied for determination of KT ions in urine and pharmaceutical preparations using potentiometric determination, standard addition and the calibration curve methods. The standard deviation computed on the results indicated excellent repeatability of the measurements. Overall, it showed satisfactory results with excellent percentage recovery comparable to and sometimes better than those obtained by other routine methods for the assay.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Ion-Exchange Chromatography
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...