Wavelength-Filter Based Spectral Calibrated Wave number - Linearization in 1.3 mm Spectral Domain Optical Coherence
Ruchire Eranga Henry Wijeisnghe1, Nam Hyun Cho1, Kibeom Park1
1Department of Electronics Engineering, Kyungpook National university, Daegu, South Korea.
Abstract:
In this study, we demonstrate the enhanced spectral calibration method for 1.3 μm spectral-domain optical coherence tomography (SD-OCT). The calibration method using wavelength-filter simplifies the SD-OCT system, and also the axial resolution and the entire speed of the OCT system can be dramatically improved as well. An externally connected wavelength-filter is utilized to obtain the information of the wavenumber and the pixel position. During the calibration process the wavelength-filter is placed after a broadband source by connecting through an optical circulator. The filtered spectrum with a narrow line width of 0.5 nm is detected by using a line-scan camera. The method does not require a filter or a software recalibration algorithm for imaging as it simply resamples the OCT signal from the detector array without employing rescaling or interpolation methods. One of the main drawbacks of SD-OCT is the broadened point spread functions (PSFs) with increasing imaging depth can be compensated by increasing the wavenumber-linearization order. The sensitivity of our system was measured at 99.8 dB at an imaging depth of 2.1 mm compared with the uncompensated case.
Related Concept Videos
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to...
UV–Vis Spectroscopy: Beer–Lambert Law
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...