Multicentre comparison Of shock efficacy using single-vs. Dual-coil lead systems and Anodal vs. cathodaL polarITY
Maria Stella Baccillieri1, Gianni Gasparini, Luca Benacchio
1Cardiovascular Division, P. Cosma Hospital, Camposampiero, Italy, hhbacci@alice.it.
Purpose:
An optimal active-can lead configuration during implantable cardioverter defibrillator (ICD) placement is important to obtain an adequate defibrillation safety margin. The purpose of this multicenter study was to evaluate the rate of the first shock success at defibrillation testing according to the type of lead implant (single vs. dual coil) and shock polarity (cathodal and anodal) in a large series of consecutive patients who received transvenous ICDs.
Methods:
This was a multicenter study enrolling 469 consecutive patients. Single- versus dual-coil leads and cathodal versus anodal polarity were evaluated at defibrillation testing. In all cases, the value of the energy for the first shock was set to 20 J less than the maximum energy deliverable from the device.
Results:
A total of 469 patients underwent defibrillation testing: 158 (34 %) had dual-coil and 311 (66 %) had single-coil lead systems configuration, 254 (54 %) received anodal shock and 215 (46 %) received cathodal shock. In 35 (7.4 %) patients, the shock was unsuccessful. No significant differences in the outcome of defibrillation testing using single- versus dual-coil lead were observed but the multivariate analysis showed an increased risk of shock failure using cathodal shock polarity (OR 2.37, 95 % CI 1.12-5.03).
Conclusions:
Both single- and dual-coil transvenous ICD lead systems were associated with high rates of successful ICD implantation, and we found no significant differences in ventricular arrhythmias interruption between the two ICD lead systems configuration. Instead, anodal defibrillation was more likely to be successful than cathodal defibrillation.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use
Cardiac Catheterization III: Left Heart Catheterization
Cardiomyopathy V: Interprofessional Care