Non-destructive vacuum decay method for pre-filled syringe closure integrity testing compared with dye ingress
Andrea Simonetti1, Filippo Amari2
1Senior Manager, Strategic Initiatives, Bonfiglioli Engineering S.r.l., TASI Group, Via Rondona, 33 - 44049 Vigarano Pieve (FE), Italy; Telephone +39 0532 715631, a.simonetti@bonfiglioliengineering.com; a.simonetti@bonfiglioliengineering.com.
The vacuum decay method is the most effective for detecting defects in prefilled syringes, offering superior reliability and repeatability compared to other container closure integrity testing methods. This ensures sterile drug product quality and safety.
Area of Science:
- Pharmaceutical Technology
- Analytical Chemistry
- Regulatory Science
Background:
- Traditional sterility testing has limitations, prompting regulatory bodies like the FDA to encourage validated physical methods for container closure integrity testing (CCIT).
- Ensuring container closure integrity is critical for maintaining sterile drug product quality, safety, and shelf-life.
- Advancements in CCIT methods and equipment are driven by regulatory, quality, and safety expectations for sterile drug products.
Purpose of the Study:
- To evaluate the capability of four different CCIT methods in detecting simulated defects in glass prefilled syringes.
- To compare the performance of vacuum decay, pharmacopoeial dye ingress, Novartis specific dye ingress, and high-voltage leak detection methods.
- To identify the most effective, reliable, and repeatable CCIT method for prefilled syringe applications.
Main Methods:
- Development and validation of four CCIT methods: vacuum decay, pharmacopoeial dye ingress, Novartis specific dye ingress, and high-voltage leak detection.
- Creation of reference prefilled syringes with known defects (5 μm, 10 μm, 20 μm) using laser-drilled holes and embedded capillary tubes.
- Testing of syringes filled with sterile water and a flu vaccine (Agrippal) using validated commercial production equipment.
Main Results:
- The vacuum decay method demonstrated the highest performance in detecting defects, showing superior sensitivity, reliability, and repeatability.
- All tested methods were evaluated for their detection rates and false result percentages across different defect types and sizes.
- An innovative solution to prevent prefilled syringe plunger movement during CCIT was presented.
Conclusions:
- Vacuum decay is the most effective non-destructive method for ensuring prefilled syringe container closure integrity.
- This method reliably detects defects regardless of the fill substance (sterile water, drug product, or air).
- The findings support the adoption of advanced CCIT methods to meet stringent regulatory requirements for sterile drug products.