Real-time automated thickness measurement of the in vivo human tympanic membrane using optical coherence tomography

Zita Hubler1, Nathan D Shemonski1, Ryan L Shelton1

  • 11 Department of Bioengineering, 2 Beckman Institute for Advanced Science and Technology, 3 Department of Electrical and Computer Engineering, 4 Department of Internal Medicine, University of Illinois at Urbana-Champaign, Urbana, IL, USA.

Abstract

Insights

Optical coherence tomography (OCT) offers a new quantitative method for diagnosing middle ear infections (otitis media). This technology accurately measures tympanic membrane thickness in real-time, improving diagnosis and monitoring of ear conditions.

Area of Science:

  • Biomedical Engineering
  • Ophthalmology
  • Medical Imaging

Background:

  • Otitis media (OM) is a common pediatric ear infection requiring accurate diagnosis.
  • Current diagnostic methods like otoscopy are largely qualitative.
  • Chronic OM may necessitate surgical intervention, highlighting the need for precise monitoring.

Purpose of the Study:

  • To develop and validate an optical coherence tomography (OCT) system for high-resolution imaging of the tympanic membrane (TM) and middle ear.
  • To create a novel algorithm for automatic, real-time measurement of TM thickness for OM diagnosis.
  • To quantitatively assess TM thickness and detect middle ear biofilms.

Main Methods:

  • An OCT system was developed for cross-sectional imaging of the TM.
  • A novel segmentation algorithm using Hough transform was created for real-time TM thickness measurement.
  • The system was tested on phantoms and in vivo human subjects.

Main Results:

  • The OCT system demonstrated high accuracy with minimal variation on phantoms.
  • Real-time in vivo imaging showed spatial and case-specific differences in TM thickness.
  • The algorithm accurately measured TM thickness in normal, acute, and chronic OM cases.

Conclusions:

  • OCT combined with the novel segmentation algorithm provides a robust, quantitative method for diagnosing OM.
  • This technology enables accurate, real-time, in vivo assessment of TM thickness.
  • The system aids in the diagnosis and monitoring of various middle ear conditions.