Related Experiment Video
Updated: Apr 17, 2026

Predicting Amputation using Local Circulating Mononuclear Progenitor Cells in Angioplasty-treated Patients with Critical Limb Ischemia
Published on: September 22, 2020
Early experimental and clinical experience with a focal implant for lower extremity post-angioplasty dissection
Peter A Schneider1, Robert Giasolli2, Adrian Ebner3
1Kaiser Foundation Hospital, Honolulu, Hawaii.
Objectives:
This study provides preliminary data on the safety and feasibility of the use of a novel focal implant for managing post-percutaneous transluminal balloon angioplasty (post-PTA) dissection.
Background:
Post-PTA dissection of the lower extremity arteries is managed with stent placement. This provides an acceptable post-intervention result but has long-term disadvantages, such as in-stent restenosis. Focal treatment of post-PTA dissection and avoidance of stents are the objectives of the Tack-It (Intact Vascular, Inc., Wayne, Pennsylvania) device.
Methods:
A preclinical study and first-in-human data are presented. Seven swine underwent superficial femoral artery device placement, with a self-expanding nitinol stent on 1 side and a series of 4 Tack-It devices on the other side. Specimens were harvested at 28 days. The clinical study included 15 limbs that underwent revascularization for critical limb ischemia (n = 9) or claudication (n = 6). Twenty-five lesions were treated in the superficial femoral (n = 8), popliteal (n = 7), and tibial (n = 10) arteries.
Results:
The preclinical study demonstrated a reduction in stenosis with the Tack-It (16.8 ± 2.6%) compared with stents (46.4 ± 9.8%). Neointimal thickness and injury score decreased with the Tack-It. Clinically, Tack-It placement resulted in acute technical success with resolution of the post-PTA dissection in 100% of lesions. There were no device-related complications or major amputations. Eighteen of the 25 lesions were available for angiographic follow-up at 1-year, and patency was 83.3%.
Conclusions:
Preclinical data suggest that the Tack-It device causes minimal vessel injury. Clinical use of the Tack-It to manage post-PTA dissection was safe and feasible in this early study and resulted in apposition of dissection flaps without stent placement.
More Related Videos
07:48Assessing Therapeutic Angiogenesis in a Murine Model of Hindlimb Ischemia
Published on: June 8, 2019
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Peripheral Artery Disease III: Interprofessional Care
Peripheral Artery Disease V: Postoperative Nursing Management