In vivo temperature rise in anesthetized human pulp during exposure to a polywave LED light curing unit
Patrício Runnacles1, Cesar Augusto Galvão Arrais1, Marcia Thais Pochapski1
1Department of Restorative Dentistry, State University of Ponta Grossa, Rua General Carlos Cavalcanti, 4748, Uvaranas, Ponta Grossa, PR 84030-900, Brazil.
Objectives:
This in vivo study evaluated pulp temperature (PT) rise in human premolars during exposure to a light curing unit (LCU) using selected exposure modes (EMs).
Methods:
After local Ethics Committee approval, intact first upper premolars, requiring extraction for orthodontic reasons, from 8 volunteers, received infiltrative and intraligamental anesthesia. The teeth (n=15) were isolated using rubber dam and a minute pulp exposure was attained. A sterile probe from a wireless, NIST-traceable, temperature acquisition system was inserted directly into the coronal pulp chamber, and real time PT (°C) was continuously monitored while the buccal surface was exposed to polywave light from a LED LCU (Bluephase 20i, Ivoclar Vivadent) using selected EMs allowing a 7-min span between each exposure: 10-s either in low (10-s/L) or high (10-s/H); 5-s-turbo (5-s/T); and 60-s-high (60-s/H) intensities. Peak PT values and PT increases from baseline (ΔT) after exposure were subjected to one-way, repeated measures ANOVAs, and Bonferroni's post hoc tests (α=0.05). Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT.
Results:
All EMs produced higher peak PT than the baseline temperature (p<0.001). The 60-s/H mode generated the highest peak PT and ΔT (p<0.001), with some teeth exhibiting ΔT higher than 5.5°C. A significant, positive relationship between applied radiant exposure and ΔT (r(2)=0.916; p<0.001) was noted.
Significance:
Exposing intact, in vivo anesthetized human upper premolars to a polywave LED LCU increases PT, and depending on EM and the tooth, PT increase can be higher than the critical ΔT, thought to be associated with pulpal necrosis.