Related Experiment Video
Updated: Apr 16, 2026

A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
Optimizing smart pump technology by increasing critical safety alerts and reducing clinically insignificant alerts
Jennifer Mansfield1, Steven Jarrett2
1Smart Pump Liaison, Carolinas HealthCare System , Charlotte, North Carolina.
Reducing clinically insignificant alerts from intravenous (IV) infusion pumps through data analysis and library limit revisions can decrease alert fatigue and enhance patient safety. This strategy improves the effectiveness of smart infusion pumps.
Area of Science:
- Medical Informatics
- Patient Safety Engineering
- Clinical Alert Systems
Background:
- Intravenous (IV) infusion pump alerts are critical for preventing harm but can be ignored due to frequent, clinically insignificant notifications.
- This desensitization to alerts poses a significant risk to patient safety.
- Limited research exists on visual alert desensitization in healthcare settings.
Purpose of the Study:
- To describe a method for reducing non-clinically relevant alerts from IV infusion pumps.
- To identify specific medications contributing to alert fatigue.
- To enhance the effectiveness of smart infusion pump safety software.
Main Methods:
- Analysis of IV infusion pump alert data using scatter plots.
- Identification of medications associated with the highest and lowest frequency of clinically meaningful alerts.
- Revision of drug library limits for specific medications based on alert data.
Main Results:
- A decrease in the number of non-clinically relevant infusion pump alerts was achieved.
- Identification of specific drug library parameters that contributed to excessive alerts.
- The process enabled targeted adjustments to reduce alert burden.
Conclusions:
- Revising drug library limits for specific medications effectively reduced clinically insignificant alerts.
- Reduced alert fatigue enhances the reliability and effectiveness of smart infusion pumps.
- This intervention provides an additional layer of safety for patient care.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Impact of Pharmacokinetic–Pharmacodynamic Models: Regulatory Decisions
Bioreactor Controls-I
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Heart Failure VI: Adjunct Therapies
Pharmaceutical Poisoning: Potential Scenarios