Related Experiment Video
Updated: Apr 16, 2026

A Practical and Novel Method to Extract Genomic DNA from Blood Collection Kits for Plasma Protein Preservation
Published on: May 18, 2013
Relevance of EDTA carryover during blood collection
Background:
The order of draw is regarded as a preanalytical issue to prevent carryover of additives during blood collection. Our objective was to prove the theory of ethylenediaminetetraacetic acid (EDTA) carryover for a closed vacuum system and the influence of EDTA on concentrations of selected biomarkers.
Methods:
To test the carryover of EDTA, a blood collection with tripotassium EDTA (K3EDTA) and subsequent non-additive tubes was simulated using distilled water as substitute for blood. EDTA concentrations were measured by tandem mass spectrometry. Then we added increasing concentrations of EDTA to heparinized blood and measured routine biomarkers, thereby simulating a carryover of EDTA whole blood and pure EDTA, respectively. Additionally, we tested for EDTA contamination and biomarker alteration in samples collected from 10 healthy volunteers by a syringe with subsequent transfer into sample tubes.
Results:
No EDTA contamination was detected in samples collected subsequent to a K3EDTA tube when adhering to guidelines of blood sampling. Magnesium, calcium, and potassium levels were altered by artificial K3EDTA whole-blood contamination as well as when adding 1 μL pure K3EDTA. Iron values were altered at EDTA concentrations of 4.4 mmol/L. All other parameters remained unaffected. A slight EDTA carryover was observed in syringe collection and subsequent transfer into EDTA and heparin tubes, however, without any biomarker alteration.
Conclusions:
An EDTA carryover during blood collection using a closed vacuum system is highly unlikely. Even if carryover of EDTA whole blood occurs, an absolute volume larger than 10 μL would be necessary to alter test results. However, contamination of samples with preloaded pure K3EDTA solution by severe neglect of current recommendations in blood collection may significantly alter testing results.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Auxiliary Complexing Reagents
Complexometric EDTA Titration Curves
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...
EDTA: Chemistry and Properties
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...